The ELx continues our tradition of offering the tightest specifications of any microplate reader. Visit BioTek to learn more about our precision microplate. ELx Operator’s Manual. Revision History. Rev Date Changes. A. 1/96 First Release. B. 5/96 Revised reader specifications. Added Scanning. View and Download Biotek ELx operator’s manual online. Absorbance Microplate Reader. ELx Laboratory Equipment pdf manual download. Also for.

Author: Shaktimuro Dom
Country: Liberia
Language: English (Spanish)
Genre: Politics
Published (Last): 2 September 2004
Pages: 360
PDF File Size: 5.78 Mb
ePub File Size: 8.79 Mb
ISBN: 876-5-39557-714-2
Downloads: 76228
Price: Free* [*Free Regsitration Required]
Uploader: Gosar

Every ELx reader is fully tested at BioTek prior to shipment and should operate properly upon initial setup.

BIOTEK ELX OPERATOR’S MANUAL Pdf Download.

The top line of the display updates to reflect this selection. Operation BioTek Instruments, Inc. Roller surface and hook-in bracket BioTek Instruments, Inc.

Dual wavelength readings can significantly reduce optical interference caused by scratched or fingerprinted microplates since the scratches or fingerprints reduce the amount of light on both wavelengths.

Biotek ELx808 Operator’s Manual: Error Codes (general)

The bulb should be replaced at this time PNand the electrical contacts in the lamp maanual cleaned. Number Of Standards To start running a reader program, press the Read key. You can start entering a starting sample identification manal from to By pressing ‘print’ button you will print only current page.

Turn off and unplug the reader. Use a clean, dry lint-free cloth to dry all wet surfaces. Don’t show me this message again.

  KARBONMONOKSIT ZEHIRLENMESI PDF

By pressing ‘print’ button you will print only current page. Contact Information outside the U.

Intended Use Statement Remain aware of the total number of controls, standards, and blanks that were originally mapped in each individual assay while mapping for the Panel assay. Instructions For Programming A New Assay The kit instructions are provided so that users can see lex808 it is possible to translate the kit wording into an ELx assay program.

Microplate Readers: ELx Absorbance Microplate Reader – Overview

Older-Style Module A failed fuse is usually an indication of another problem that a new fuse is not likely to remedy. For assays defined onboard by a user Assay Assay55it is Standardized the presentation of significant digits.

Single Or Dual Wavelength Examples Calls are assigned to sample wells as follows: Got it, continue to print. The first digit will be 0, 1, 2, or A. Enter text from picture: Page Ensure that the manuxl is set flat against its housing before reinstallation in the instrument. Recommended Maintenance Schedule Designed to automatically perform endpoint and kinetic analysis, the reader can measure the optical density of solutions in well microplates between nm and nm. Warnings And Precautions Filter wheel motor did not find the home sensor; filter wheel not.

Reconfigure or replace fuses as needed see next page.

manua Another Transformation Example 0 standardand multiplies the results by To avoid system instability, make sure the printer and reader are turned OFF before connecting them. Liquid Tests Prepare Test Solutions BioTek offers a dye solution PN25 mL; ormL that may be used in the stock solution formulation for Liquid Tests 1 and 2, or, if you prefer, you may use the dye solution described below.

  HANDBOOK OF CHEMOMETRICS AND QUALIMETRICS PART A PDF

If there are any mismatches, correct them and re-run the test. Allow the lamp to cool for at least ten minutes. Manual mapping is recommended to set up the appropriate map by placing the standards, controls, and samples in only the A, C, E, and G rows.

Select the Operating Environment Page 34 – 3: When using nm filters, the analog gain must be increased.

To obtain reader software information: The mathematical computations described below may then be performed and the template kept for future data reduction. Installation Page 32 – Manula Registration Page 33 – 2: Usa Nanual Class A Modified appearance of display screens. Part 1, General requirements.

Biotek ELx808 Manuals

Not all features of the software discussed in this chapter are available on custom instruments. Newer-style Module As shown below, a small voltage selector card is located to the right of the fuse holder. The Startup Screen Chapter 3: This manual also for: The Startup Screen Terms Used in the Error Codes Tables Table Of Contents Before Repackaging The Instrument